When a residue screen informs lot acceptance, collect the sample under the receiving plan and retain the link between the sample and the shipment. Match the assay to the specific drug and food type before preparing the sample. Clear traceability matters most when a shipment contains several lots or a supplier sends both raw and processed material.
Set up the bench around the reading window
Lay out the cards, timer, sample IDs and preparation tools in the order of testing. Start a group only when the operator can apply the sample and read every card within its method window. Use the assay’s control indicator to determine whether each result is valid. Read the designated reaction area according to the card instructions, since different assay formats can assign different meanings to line intensity.
Review an unexpected result
Check card storage, expiry, the control indicator, sample identity and the recorded application and read times. For a milk sample, also check that the preparation and test card are specified for that milk type. Store the initial result alongside any repeat. Where the quality procedure calls for confirmation, retain the right sample portion and refer it through the prescribed process.
Release a useful test record
Report the target, matrix, assay, sample lot and valid outcome together. For a portable HM-SYJC workflow, group tests by compatible assay and schedule the preparation and reading windows before starting the shift.
Shandong Hengmei Electronic Technology Co., Ltd.
Planning an HM-SYJC workflow
Prepare a bench area for the HM-SYJC reader and a separate area for handling incoming samples. Keep assay lots and timing visible at the reading station. Record the control result with each shipment identifier so a rejected reaction can be traced.
Control timing and preserve the primary result
Colour development in a rapid-test card changes over time. Log the time of application and the time of reading and assign a manageable number of cards to each operator. When the control indicator does not meet the kit instructions, record the result as invalid and investigate handling or reagent condition. A valid presumptive positive is a different outcome. Retain the original card reading, the card lot and a traceable sample portion for the required confirmation. If the exact target drug and matrix have different extraction steps, use separate worksheets even when a single reader is used.
Check the calculation before reporting
36 lots × 2 target assays = 72 sample reactions before controls or repeats. If 6 reactions are repeated and 4 controls are required, reserve 82 reactions for the shift. Calculate consumable cost from reactions used and divide by the number of lots that actually receive a report.
Troubleshooting and result review
Why can the same drug give different results in milk and meat?
Matrix composition and sample preparation affect the assay response. Use the card and extraction procedure validated for the material being tested.
Keep the original result and control readings with the repeat. Record the cause investigated and the disposition of the lot so the next reviewer can reconstruct the decision.
Prepare the batch and establish the decision basis
Before measuring veterinary drug residues, review the incoming list for milk, raw meat and seafood receiving lots. Match every item to the approved method and assign an identifier that stays with the prepared portion through transfer and measurement. Agree on the drug-specific screening level and the laboratory's confirmation policy and confirm who may release the result. Prepare a tray or worklist that distinguishes routine samples, controls and repeats. A complete batch plan also tells the operator which specimens can be run together and where the sample or original package must be retained for later review.
Set out matched negative and fortified material for each drug and matrix and keep controls identifiable through preparation. Confirm the instrument programme, current reagent or reference material identity and the test conditions before loading samples. Record card storage, incubation and read time. With a time-dependent procedure, start each group only when the operator can complete reading within the defined window. Keep an observation field for changes during preparation, such as unusual sample appearance, transfer problems or a departure from the planned handling conditions.
Follow the sample through preparation and measurement
| Check | Evidence to retain |
|---|---|
| Before preparation | Confirm sample identity, matrix and suitability for the drug-specific screening level and the laboratory's confirmation policy. |
| During preparation | Preserve the sample ID and record card storage, incubation and read time. |
| Before interpretation | Review matched negative and fortified material for each drug and matrix first and identify a control indicator outside the kit acceptance criteria. |
| At report review | Check the method, original reading, result basis and any follow-up prompted by cross-reactivity or a matrix response that requires confirmation. |
When an aliquot is prepared or transferred, record the receiving container or position. If two results need comparison, check whether they came from two independently prepared portions or two readings of the same prepared portion. These observations help locate variation at different stages of the work. Keep original values and preserve any separate sample mass, volume, units and correction factors. If the instrument already applies a factor automatically, record that setting with the reading so the reviewer can reproduce the reported value.
Decide whether to stop, repeat or confirm
When you see a control indicator outside the kit acceptance criteria, pause the affected result and assess the controls and handling first. Review the timing and operator notes before adding another measurement. If the problem is systematic, such as reagent condition or a failed preparation control, identify all samples affected since the last acceptable check. Follow the batch procedure for recovery and retain the original results. When an individual specimen is unsuitable, record why the specimen was rejected or resampled, and keep it tied to the incoming lot record.
To investigate cross-reactivity or a matrix response that requires confirmation, review sample preparation and instrument conditions in order. A repeat of the same prepared portion can help assess repeatability of the reading. A fresh preparation can explore weighing, transfer or preparation variation. Select the type of repeat according to the problem and document the reason before starting. Where the decision requires a separate confirmatory method, prepare the retained portion for that method and preserve identification through the transfer. A valid unexpected result should remain visible through investigation and disposition.
Release a report that can be reviewed later
Write the result with the sample and lot identity, the method and its relevant units, and any exception or repeat decision. Report card storage, incubation and read time in the worksheet when those details affect interpretation. Retain the control response and associate it with the samples that shared it. The report should enable a supervisor to determine whether the method covered the material examined and whether the applicable decision criteria were met. Use consistent units and record how any calculation was applied before comparing a result with a purchase or process specification.
For repeated batches, review trends in assay cards, extraction supplies and confirmatory laboratory charges along with the rate of invalid tests, repeat work and delayed decisions. This can identify whether delays arise from preparation, measurement, consumables or report review. Train another operator to reconstruct a past batch from the original information and investigate a control indicator outside the kit acceptance criteria. Where a step cannot be reconstructed, improve the record at the point where the work takes place. The resulting worksheet is useful for training, follow-up with suppliers and review of process changes.